> Measuring and Control > TOPAX® Controllers > Controller TOPAX® LF 2
 |
| Select from the files and links below: |
|
General DescriptionTOPAX® LF 2 Controller- Conductivity parameter measuring and controller device (conductive conductivity)
- Temperature controller device.
Measuring unit - Measures: Conductivity (conductive measuring principle)
- Metering range:
- 0 ... 2,000 μS/cm C = 0,05/cm - 0 ... 20,00 μS/cm C = 0,05/cm - 0 ... 200,0 μS/cm C = 0,05/cm - 0 ... 2,000 mS/cm C = 0,2/cm
- Temperature metering range: -22… +284 °F
- Display: Measurement values with relative units. Status messages, sensors. Calibration, controller and alarm.
- Temperature compensation: Either manual or automatic, Pt 100 measuring principle.
- Temperature coefficient: 32 - 46.4% / °F
Controller unit - Switch points: 2 switch points with adjustable effective direction setting
- Controller options ON/OFF controller, with optional hysteresis
- Proportional (P) controller - pulse-pause, pulse- frequency or continuous controller. Proportional integral (PI) controller - pulse-pause, pulse- frequency or continuous controller
- Hysteresis: freely adjustable within the metering range. Proportional gain KP: freely adjustable within the metering range.
- Reset time TN: 0 - 2000 sec
- Minimum impulse: 0.1 - 9.9 sec
- Pulse + pause time: 02 - 99 sec
- Impulse frequency: 100 - 7200 impulses/h
- Switch-on delay: 0 - 200 sec
- Alarm function: minimum and maximum threshold value and delay time
|
 |